PET Daily QA Instructional Video with Step-by-Step

MRIPETCTSOURCE has provided a detailed step-by-step instructional video. The PET Daily QA procedure should be performed on a daily basis prior to commencing patient scans. Follow along with detailed instructions and video provided below. We are always adding new resources, subscribe to receive a notification every time we post relevant content.

PET Daily QA Procedure Video

Link provided below to step-by-step procedure to Establishing Baseline once current reading has been analyzed and all field pass.

PET Daily QA Step-by-Step Instructions

Time needed: 15 minutes

How-to: PET Daily QA procedure- Take Current Reading Part 1

  1. Select Exam RX on right-side monitor.

  2. Click on [Daily Prep] on left side monitor.

  3. Select [PET Daily QA] on upper right corner of left-side monitor.

    Tabs will be displayed across the top of your screen with following selections:
    VIEW BASELINE: Displays the Graph for Baseline Reading and disables the [ESTABLISH A BASELINE] button. The button label changes to VIEW CURRENT. The Graph for Baseline Reading represents the currently established Daily QA baseline, if one exists.
    VIEW CURRENT: Displays the Graph for Current Reading and activates the [ESTABLISH A BASELINE] button. The button label changes to VIEW RATIO. The Graph for Current Reading displays the most recently acquired Daily QA data. When you click the [ESTABLISH A BASELINE] button, the system copies the contents of the selected values over the corresponding contents of the Graph for Baseline Reading file. So, you may update a single value, multiple values or all the values.
    VIEW RATIO: Displays the Graph for Ratio Data and disables the [ESTABLISH A BASELINE] button. The button label changes to VIEW BASELINE.

  4. Select [Take Current Reading].

  5. A dialogue box opens to display the message: “The Transmission Source will be Exposed! Proceed?”

  6. Select [OK].

  7. Press the flashing “Move to Scan” (if required) then press start scan buttons to initiate the acquisition.

  8. The software automatically processes the data, then opens an Image Quality and Graph GUI with details of pass or fail status. The GUI displays selection buttons along the top of the panel, and a table of values at the bottom of the GUI panel.

  9. Move the cursor over the strip charts to display the Module, Block, Crystal Row, Crystal Column, and Data Value values along the bottom of the screen. Follow the procedures in the service documentation to troubleshoot and/or replace any bad modules, and recalibrate the system.

  10. When satisfied with the Image Quality and Graph display results, proceed to Establish a Baseline, and set the Baseline Date.

  11. Process is Complete. You may now click Exit to close QA panel

  12. Click [YES] to close the dialogue box

  13. Establish Baseline

    From the pull down menus select [View Current Reading] and [View Data] to display the Graph for Current Reading panel.
    -If the graph contents look acceptable, click [Establish a Baseline], to open a list of prior Baseline Data.
    -If the graph contents are not acceptable, troubleshoot the problem

  14. Close data table.

    Use the mouse to move the cursor over a box or label, then click-left to display a check mark in the box. Click a second time to toggle the check mark off. The data table at the bottom of the panel disappears.

  15. Write New Data.

    Select all the Data Types, then click [Accept] to overwrite the baseline information with the corresponding check marked data.

  16. Exit Window.

    Click [Exit] to close the Baseline Data Selector panel. Detector modules in the detector ring may be made by different manufacturers which can result in shade variations from module to module in the graphic display. These variations are not “Stand alone” indicators that the module is out of specification or impact image quality. Always refer to the data specified and consult with service engineer if any issues arise.

  17. Process Complete

    Click [EXIT] to close the Daily QA panel.

Tab Selections

  • VIEW BASELINE: Displays the Graph for Baseline Reading and disables the [ESTABLISH A BASELINE] button. The button label changes to VIEW CURRENT. The Graph for Baseline Reading represents the currently established Daily QA baseline, if one exists.
  • VIEW CURRENT: Displays the Graph for Current Reading and activates the [ESTABLISH A BASELINE] button. The button label changes to VIEW RATIO. The Graph for Current Reading displays the most recently acquired Daily QA data. When you click the [ESTABLISH A BASELINE] button, the system copies the contents of the selected values over the corresponding contents of the Graph for Baseline Reading file. So, you may update a single value, multiple values or all the values.
  • VIEW RATIO: Displays the Graph for Ratio Data and disables the [ESTABLISH A BASELINE] button. The button label changes to VIEW BASELINE.

Key Takeaways

The GE PETCT Daily QA is a simple process that should be performed on a routine basis for optimal image quality. Refer to the model-specific operator manual for your system for additional details.

Providing the Most Powerful Resources

We offer world-class service for your MRI, CT, and PET systems. OEM-trained service engineers with years of experience provide expert service and technical support anywhere in the world. Call us today to schedule your first service appointment. More information is listed on our Service and Support page.

We have a helpful library of resources available and are always adding new content. Check back often for new articles, guides, and manuals!

Contact Us For All of Your Imaging Needs

We bring you the highest quality products and services at the most competitive price. The diagnostic imaging market is flooded with requests for parts that are in high demand, hard to find or are no longer manufactured. We outshine our competitors by finding the parts you need and consistently delivering expedited service for the highest quality products at the lowest price. Contact us today!

Subscribe To Our Newsletter To Stay Up To Date With The Latest News, Exclusive Offers, And Giveaways!

The information provided by Medical Imaging Source (“we,” “us,” or “our”) on https://www.medicalimagingsource.com (the “Site”) is for general informational purposes only. All information on the Site is provided in good faith, however we make no representation or warranty of any kind, express or implied, regarding the accuracy, adequacy, validity, reliability, availability, or completeness of any information on the Site. UNDER NO CIRCUMSTANCE SHALL WE HAVE ANY LABILITY TO YOU FOR ANY LOSS OR DAMAGE OF ANY KIND INCURRED AS A RESULT OF THE USE OF THE SITE OR RELIANCE ON ANY INFORMATION PROVIDED ON THE SITE. YOUR USE OF THE SITE AND YOUR RELIANCE ON ANY INFORMATION ON THE SITE IS SOLELY AT YOUR OWN RISK.

Scroll to Top