Medcam AP board replacement procedure for the GE Signa LX 1.5T MRI system. This resource is part of our growing library of GE MRI service manual procedures.
All GE MRI service methods documents are available for download in the GE service documentation library.
How To Replace AP Board (GE Signa LX 1.5T MRI)
Time needed: 15 minutes
Medcam AP Board Replacement (GE Signa LX 1.5T MRI)
- Remove front cover from GE Signa LX Systems Cabinet.
- Turn off main circuit breaker, CB1.
- Remove top and bottom screw on the front of Medcam AP board.
- Remove Medcam AP board from ISE/TPS Chassis
- Install New Medcam AP Board
- Turn on main circuit breaker, CB1
- Install Systems Cabinet Covers
- Perform TPS Reset.
- Perform Medcam AP Board Functional Checks
Conclusion
MRIPETCTSOURCE provides the best radiology resources for imaging professionals. Our growing library of helpful guides is available 24/7 to assist you in resolving MRI issues. Subscribe to our newsletter above to stay up to date on the latest releases.
Click here to view the GE MRI Service Methods Video Playlist on the MRIPETCTSOURCE YouTube Channel or check out some of the related pages below.
Join the Medical Imaging Source Community!
Subscribe To Our Newsletter To Stay Up To Date With The Latest News, Exclusive Offers, And Giveaways!
Free Cancer Patient Care Package, Gifts and More!
In this article, our primary focus shifts to what matters most: the patients within our community. We want you to […]
MRI (Magnetic Resonance Imaging)
What Is MRI? Magnetic resonance imaging is a medical imaging marvel that produces a symphony of sounds to produce high […]
MRI Cold Head Failure FAQ
What causes MRI cold head failure? In this article, we reveal expert tips to keeping your MRI cold head running […]
Healthcare Cyber Attack Survival Guide
Change Healthcare 2024 data breach affected over 100 million people, as confirmed by the HHS in an October 2024 Cybersecurity Incident Update.
15 Tips to Reduce Patient No-Show Rates
High patient non-completion rates, encompassing no-shows and patients refusals, significantly impact profitability and operational efficiency for imaging centers. These factors […]
The information provided by MRIPETCTSOURCE (“we,” “us,” or “our”) on https://www.medicalimagingsource.com (the “Site”) is for general informational purposes only. All information on the Site is provided in good faith, however we make no representation or warranty of any kind, express or implied, regarding the accuracy, adequacy, validity, reliability, availability, or completeness of any information on the Site. UNDER NO CIRCUMSTANCE SHALL WE HAVE ANY LABILITY TO YOU FOR ANY LOSS OR DAMAGE OF ANY KIND INCURRED AS A RESULT OF THE USE OF THE SITE OR RELIANCE ON ANY INFORMATION PROVIDED ON THE SITE. YOUR USE OF THE SITE AND YOUR RELIANCE ON ANY INFORMATION ON THE SITE IS SOLELY AT YOUR OWN RISK.